技术参数
优异的移液舒适性!
Gilson执行的规格标准比ISO管理机构的要求更加严格,允许的误差范围比ISO8655规定的低一倍。
我们的移液器生产工厂拥有认证的校准实验室来控制每一支Gilson移液器的精确度和准确度。
同时,每支移液器会单独校准,交付时随附性能检验报告。每支移液器机身有雕刻的序列号以便追溯。
容量规格
型号 | 订货号 | 量程(μL) | Gilson允许的最大误差 | ISO 8655允许最大误差 | |||
系统误差(μL) | 随机误差(μL) | 系统误差(μL) | 随机误差(μL) | ||||
P2G
|
F144054P F144054M |
0.2 | ±0.024 | ≤0.012 | ±0.08 | ≤0.04 | |
0.5 | ±0.025 | ≤0.012 | ±0.08 | ≤0.04 | |||
2 | ±0.030 | ≤0.014 | ±0.08 | ≤0.04 | |||
P10G
|
F144055P F144055M |
1 | ±0.025 | ≤0.012 | ±0.12 | ≤0.08 | |
5 | ±0.075 | ≤0.030 | ±0.12 | ≤0.08 | |||
10 | ±0.100 | ≤0.040 | ±0.12 | ≤0.08 | |||
P20G
|
F144056P F144056M |
2 | ±0.10 | ≤0.03 | ±0.20 | ≤0.10 | |
10 | ±0.10 | ≤0.05 | ±0.20 | ≤0.10 | |||
20 | ±0.20 | ≤0.06 | ±0.20 | ≤0.10 | |||
P100G
|
F144057P F144057M |
10 | ±0.35 | ≤0.10 | ±0.80 | ≤0.30 | |
50 | ±0.40 | ≤0.12 | ±0.80 | ≤0.30 | |||
100 | ±0.80 | ≤0.15 | ±0.80 | ≤0.30 | |||
P200G
|
F144058P F144058M |
20 | ±0.50 | ≤0.20 | ±1.60 | ≤0.60 | |
100 | ±0.80 | ≤0.25 | ±1.60 | ≤0.60 | |||
200 | ±1.60 | ≤0.30 | ±1.60 | ≤0.60 | |||
P1000G
|
F144059P F144059M |
100 | ±3 | ≤0.6 | ±8 | ≤3.0 | |
500 | ±4 | ≤1.0 | ±8 | ≤3.0 | |||
1000 | ±8 | ≤1.5 | ±8 | ≤3.0 | |||
P5000G
|
F144066 | 500 | ±12 | ≤3 | ±40 | ≤15 | |
2500 | ±15 | ≤5 | ±40 | ≤15 | |||
5000 | ±30 | ≤8 | ±40 | ≤15 | |||
P10mL G
|
F144067 | 1000 | ±30 | ≤6 | ±60 | ≤30 | |
5000 | ±40 | ≤10 | ±60 | ≤30 | |||
10000 | ±60 | ≤16 | ±60 | ≤30 |